KMID : 0372919960170010019
|
|
Journal of Biomedical Engineering Research 1996 Volume.17 No. 1 p.19 ~ p.24
|
|
High-Order Surface Gradient Coil Design Using Target Field Approach
|
|
Lee JK
Yang YJ/Jeong ST/Choi HJ/Cho ZH/Oh CH
|
|
Abstract
|
|
|
The purpose of this paper is to design high-order (or radial) surface gradient coil (SGC), which can provide multi-dimensional spatial selection. Although the spatial Selection with High-Order gradienT (SHOT) can provide a 2-D selection with only one selective RF pulse, the high-order gradient pro- duced by conventional cylindrical-shape coils has not been clinically useful due to the large selection size caused by the limited radial gradient intensity. However, by using the proposed high-order SGCs located near the imaging region, the size of volume selection can be reduced to a clinically useflll size of 1-2 cm in diameter by applying stronger radial gradient field with much less gradient driving power. So far radial SGCs have been designed by using the field component method and may cause distortion in the selection shapes. In this paper, by using the target field approach for the coil design, selected volumes became almost circular. A 40 cm-by-40 cm Z2_surface gradient coil has been designed and implemented by using the target field approach. Phantom and volunteer studies have been performed Experimental results using spatially localized MRI show good agreement to the theoretically predicted behavior.
|
|
KEYWORD
|
|
|
|
FullTexts / Linksout information
|
|
|
|
Listed journal information
|
|
|