Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0372919960170020179
Journal of Biomedical Engineering Research
1996 Volume.17 No. 2 p.179 ~ p.188
32-Channel EEG and Evoked Potential Mapping System
¾Èâ¹ü
¹Ú´ëÁØ/À¯¼±±¹/À̼ºÈÆ/ÇÔÀ±Á¤
Abstract
A clinically oriented 32-channel electroencephalogram (EEG) and evoked potential (EP) mapping system has been developed EEG and EP signals acquired from 32-channel electrodes attached on the heroid surface are amplified by a pre-amplifier which is separated from main amplifier and is located near the patient to reduce signal attenuation and noise contamination between electrodes and the amplifier. The amplified signals are further amplified by a main amplifier where various filtering and gain contr61 are achieved An automatic artifact rejection scheme is employed using neural network-based EEG and artifact classifier, by which examination time is substantially reduce4 The continuously measured EEG sigrlals are used for spectral mapping, and auditory and visual evoked potentials measured in synchronous to the auditory and visual stimuli are used for temporal evoked potential mapping. A user-friendly graphical interface based on the Microsoft Window 3.1 is developed for the operation of the system. Statistical databases for comparisons of group and individual are included to support a statistically-based diagnosis.
KEYWORD
FullTexts / Linksout information
Listed journal information