Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0372919970180030217
Journal of Biomedical Engineering Research
1997 Volume.18 No. 3 p.217 ~ p.222
Hydroxyapatite Coating on Ti Plate by a Dipping Method
Lee Jun-Hui

Kim Seok-Yeong
Kim Young-Kon
Lee In-Seop
Abstract
Hydroxyapatite(HA)-coated metal composites were made by the dipping method. The specimen substrates were Ti plates with a thickness of 2 mm. The HA coating was carried out in HA-sol for 1 min by the dipping method. The concentration of HA-sol for the coating ranged from 3.28 to 9.99 wt%. Excellent coating was observed on Ti substrate dipped once in 9.99 wt% sol. Preparation of Ti plates by sandblasting provided the better environment for coating HA on Ti surface than non-treated surface. As the concentration of sol increased, the weight change and the coating thickness increased. Above 7 wt% sol, they increased sharply.
KEYWORD
Hydroxyapatite, Dipping method, Thin coating
FullTexts / Linksout information
Listed journal information