KMID : 0372919980190040333
|
|
Journal of Biomedical Engineering Research 1998 Volume.19 No. 4 p.333 ~ p.339
|
|
Effect of Heat Treatment on Properties of Tungsten Embolization Coils
|
|
Son Ung-Hui
Park Jea-Hyung Hong Sun-Hyeong Shin Gyeong-Min
|
|
Abstract
|
|
|
The effect of heat treatment conditions on the properties of tungsten(W) embolization coils has been evaluated. The W embolization coils were fabricated by coiling the primary W coil with a diameter of 30.mm on a alumina rod mandrel with a diameter of 2.92mm. The secondary coils were then heat treated at 475and 600 for various heat treatment time ranging from 5 minutes to 36hours. The pitch distance, diameter and shape retention capability of the W embolization coils were characterized after the heat treatment. The pitch distance of the W embolization coils increased with the heat treatment time. The diameter of W emboliazation coils decreased continuously with heat treatment time. The shape retention capability of the W embolization coils increased with the heat treatment time due to an increase in elasticity by formation of tungsten oxide film on W coil surface during the heat treatment. The heat treatment condition of W embolization Col at 600 for 20 minute was considered desirable based on the optimization of the shape retention capability, pitch distance and secondary coil diameter after heat treatment.
|
|
KEYWORD
|
|
Embolization coil, Heat treatment, Pitch distance, Shape retention capability, Oxide film
|
|
FullTexts / Linksout information
|
|
|
|
Listed journal information
|
|
|