KMID : 0375319920140010081
Journal of Clinical Pathology and Quality Control
1992 Volume.14 No. 1 p.81 ~ p.86
Evaluation of IMx System (MEIA) in the Determination of TSH Concentration


Abstract
Microparticle enzyme immunoassay (MEIA) for thyrotropin(TSH) was evaluated and results were compared with those of immunoradiometric assay(IRMA) procedure.
The lower limit of detection of the assay(Mean+3SD of the zero standard) was 0.00 mIU/L. coefficient of variations of the assay(Mean+3SD of the zero standard) was 0.00mIU/L. Coefficient of variations of within day precision was 2.6-7.9% and those
of
between day precision was 6.6-11.1%. Over the range 0.00 mIU/L~37.4mIU/L, correlation between the MEIA(Y) and two site IRMA(X) was excellent (Y=1.763X-0.2362, r=0.9958). TSH concentrations measured in 828 healthy subjects were ranged from
0.4mIU/L
to
3.2mIU/L (mean 1.25mIU/L, SD 0.61).
Conclusively, MEIA is more fully automated than IRMA, well correlated with IRMA, can be reliable method for screening test of thyroid function.