Journal of Clinical Pathology and Quality Control
1993 Volume.15 No. 1 p.103 ~ p.110
Comparison of Hb Alc Micro Column Test With Imx Glycated Hemoglobin Assay


Abstract
We compared Hb Aic micro column ion exchange chromatography(Bio-Rad, U.S.A. ) with Imx Glycated hemoglobin assay(Abbott laboratories. U.S.A.) for testing of HbAlc.
The latter uses Ion Capture assay in principle, that is a boronate affinity binding assay, which measurs percent glycated hemoglobin (% GHb), and also reports standardized percent hemoglobin Alc(%Alc). Within-run coefficients of variation(CVs) of
%
Hb
Alc by Imx assay ranged from 3.09 to 5.75. the reference ranges were 5.1-8.2%(mean 6.68%)for glycated hemoglobin and 4.6-6.7%(mean 5.66%)for Hb Alc by Imx assay. The correlation study between IMx Glycated Hemoglobin Assay and Hb Aic micro column
test
resulted in % Hb Aic IMx(Y)=1.17 Bio-Rad(X)+0.36(n=40, r=0.89(P<0.05).