Journal of Clinical Pathology and Quality Control
1995 Volume.17 No. 2 p.247 ~ p.252
Evaluation of HemoCue Glucose : Portable Glucose Meter
Àü»çÀÏ
À̸íÈñ/±èÁø±Ô
Abstract
Background: We evaluated a new portable blood glucose meter, HemoCue (HemoCue AB, Angelholm, Sweden) which adopts the glucose dehydrogenase based spectrophotometry method.

Methods and Results: The within-rum imprecisions (CV) of HemoCue Glucose were ranged from 1.32% to 3.18%. The between-run imprecisions (CV) were ranged from 1.60% to 2.10%. The linearity study revealed R©÷=0.9844 (p=0.0001) in HemoCue Glucose. The correlation equations
and coefficients of four portable blood glucose meters (Y); compared with Hitachi 747 (Hitachi, Japan) as the laboratory reference (X) were Y=0.85X-12.17, R©÷=0.9328 (p=0.0001) in Accutrend (Boeringer Mannheim, Melpitas, CA), Y=1.12X-7.44, R©÷=0.9755 (p=0.0001) in Glucocard (Kyoto Daiichi Kagaku, Kyoto, japan) and Y=0.93X+7.87, R©÷=0.9773) (p=0.0001) in HemoCue Glucose, which showed best correlation among them.

Conclusions: HemoCue Glucose was satisfactory in results of within-run and between-run precisios. HemoCue Glucose also showed acceptable results linearity. HemoCue Glucose showed best correlation with Hitachi 747 among four portable blood glucose meters.
KEYWORD
Portable blood glucose meter (HemoCue Glucose), Glucose dehydrogenase