Journal of Clinical Pathology and Quality Control 1995 Volume.17 No. 2 p.247 ~ p.252
|
|
Evaluation of HemoCue Glucose : Portable Glucose Meter
|
|
Àü»çÀÏ
À̸íÈñ/±èÁø±Ô
|
|
Abstract
|
|
|
Background: We evaluated a new portable blood glucose meter, HemoCue (HemoCue AB, Angelholm, Sweden) which adopts the glucose dehydrogenase based spectrophotometry method.
Methods and Results: The within-rum imprecisions (CV) of HemoCue Glucose were ranged from 1.32% to 3.18%. The between-run imprecisions (CV) were ranged from 1.60% to 2.10%. The linearity study revealed R©÷=0.9844 (p=0.0001) in HemoCue Glucose. The correlation equations
and coefficients of four portable blood glucose meters (Y); compared with Hitachi 747 (Hitachi, Japan) as the laboratory reference (X) were Y=0.85X-12.17, R©÷=0.9328 (p=0.0001) in Accutrend (Boeringer Mannheim, Melpitas, CA), Y=1.12X-7.44, R©÷=0.9755 (p=0.0001) in Glucocard (Kyoto Daiichi Kagaku, Kyoto, japan) and Y=0.93X+7.87, R©÷=0.9773) (p=0.0001) in HemoCue Glucose, which showed best correlation among them.
Conclusions: HemoCue Glucose was satisfactory in results of within-run and between-run precisios. HemoCue Glucose also showed acceptable results linearity. HemoCue Glucose showed best correlation with Hitachi 747 among four portable blood glucose meters.
|
|
KEYWORD
|
|
Portable blood glucose meter (HemoCue Glucose), Glucose dehydrogenase
|
|