KMID : 0375319960180020295
|
|
Journal of Clinical Pathology and Quality Control 1996 Volume.18 No. 2 p.295 ~ p.304
|
|
Laboratory Evaluation of YeongDong's Urine Analyzer URISCAN S-300 and Uriscan Strip
|
|
±è¿µ±â
±è°æö/ÇÑ»ó±â/ÀÓÀμö/¹Ú¿µ³²
|
|
Abstract
|
|
|
Background:
There is no Korean products of automatic urine analyzer so far. RecentlyYeongdong pharmaceutical company has developed URISCAN S-300 employing charge-coupled-device (CCD) which is highly sensitive than light-emitting diode (LED) sensor. We
evaluated
the analytical performance of URISCAN S-300 and Uriscan Strip in the aspects of precision, accuracy, linearity and correlation with two other urine analyzers.
Methods:
We evaluated the precisions of within-run, between-run and day to day Two quality control materials (Bio Rad and Yeongdong's) and 106 patients samples were used for evaluation.
Results:
The imprecisions of within-run urinalysis were ranged 0/60 to 7/60, those of between-run precisions 0% to 5 % and those of day-to-day 0% to 10% in ketones. Linearity was good and the correlations with the foreign reagents and instruments were
very
good except specific gravity.
Conclusions:
The newly developed Yeongdong's urine analyzer URISCAN S-300 and URISCAN strip disclosed good performance.
|
|
KEYWORD
|
|
|
|