KMID : 0375320000220020255
Journal of Clinical Pathology and Quality Control
2000 Volume.22 No. 2 p.255 ~ p.258
The Evaluation of Standard 50% Hemolytic Complement Assay Using In-house Reagents
Kang Hee-Jung

Lee Eun-Hee
Abstract
Background : The complement system plays an important role in the identification and removal of foreign substances and immune complexes and in the stimulation of inflammatory response. The
standard 50% hemolytic complement (CH50) assay is the most conventional method for functional
activity of the classical complement pathway but it has been avoided in clinical laboratories because of labor intensive procedure and lack of standard protocol. This study was carried out to
set up and evaluate the conventional CH50 test using in-house reagents including sensitized sheep
erythrocytes and standard sera.

Methods : CH50 assay was performed according to the modified Mayor¢¥s method. The linearity,
within-run and between-run precision was evaluated.

Results : The within run coefficient of variation (CV) was less than 5% and between run CV was 16.7%, which was improved to 2.8% when in-house standard sera was used as calibrator in each
test. The linearity was well preserved (R2 = 0.9993).

Conclusions : CH,, test showed excellent precision and linearity with use of in-house standard sera. It would be very useful for the evaluation of classical complement pathway activity in clinical immunology labarotory.
KEYWORD
Complement activity, CH50, Mayors method, In-house standard sera