KMID : 0376319950070010229
Dental Journal of CNU
1995 Volume.7 No. 1 p.229 ~ p.238
A study on failure strength of the implant prosthesis luted with resin cement between gold cylinder and superstructure


Abstract
This study was performed to determine the failure strength and pattern of implant prosthesis luted with resin cement between gold cylinder and superstructure, and to evaluate the bonding strength of resin cement.
To evaluate failure strength and pattern, the groups were divided into 2.
Group 1: Casted gold cylinder
Group 2: Luted with resin cement between gold cylinder and superstructure.
To evaluate effects of the bond strength of the implant prosthesis luted with resin cement according to storage condition, the groups were divided into 3;
Group A: Stored in water at 37¡É for 24 hours.
Group B: 1000 cycles thermocycled between 5¡É and 55¡É in water.
Group C: Stored in water at 37¡É for 30 days.
@ES The results were as follows;
@EN 1. Failure was found at gold screw in all specimens of group1 & 2.
2. The bond strength in group 1 and 2 was 189.86 and 188.14Kgf. There was no statistically significant difference between group 1 and 2(P>0.05).
3. The bond strength of group A, B and C was 53.28, 45.86 AND 39.29Kgf. There was a statistically significant difference among the bond strength in groups A, B and C(P<0.05). The result suggest the advantage of an improved fit of superstructure
to
the
abutment and a simple procedure. But there was a measurable decrease of the bond strength according to storage condition. So, Further research is necessary to evaluate the implant prosthesis lute with resin cement between gold cylinder and
superstructure.