KMID : 0379520100260030233
|
|
Çѱ¹µ¶¼ºÇÐȸÁö 2010 Volume.26 No. 3 p.233 ~ p.236
|
|
Assay of In Vivo Chromium with a Hollow-fiber Dialysis Sensor
|
|
Ly Suw-Young
Yoo Hai-Soo Jung Min-Ki Ko Kwang-Hee Kim Byung-Jin Lee Ki-Chul Choi Byung-Min
|
|
Abstract
|
|
|
The analytical in vivo chromium ion was searched for using a voltammetric hollow-fiber dialysis sensor via square wave stripping voltammetry (SW), cyclic voltammetry (CV), and chronoamperometry. Under optimum parameters, the analytical results indicated linear working ranges of 50~400 §·/l CV and 10~80 §¶/l SW within a 30-sec accumulation time. The analytical detection limit (S/N) was 6.0 §¶/l. The developed method can be applied to in vivo tissues and in ex vivo toxicity assay, as well as to other materials that require chromium analysis.
|
|
KEYWORD
|
|
In vivo, Chromium, Voltammetriy, Hollow fiber, Dialysis sensor
|
|
FullTexts / Linksout information
|
|
|
|