Accuracy and Satisfaction with IVIC300 (Intravenous infusion controller)
Park Jung-Hee

Yang Nam-Young
Na Moon-Jun
Go Young-Jin
Kim Ki-Suk
Kim Young-Aue
Abstract
Purpose: The purpose of this study was to determine the accuracy and satisfaction with IVIC300 (Intravenous infusion controller).

Method: The subjects consisted of 90 nurses and 100 patients. Data collected during July 2013 were analyzed using descriptive statistics, t-test, ANOVA, and X2-test (Chi-Square).

Result: The error value of IVIC300 was less than that of Dosi-flow. Differences of error values according to variable of fluid infusion (needle¡¯s gauge, patient¡¯s moving, patient¡¯s calls, and nurse¡¯s arabitary calls) were not significant. The mean scores for satisfaction with IVIC300 of nurses (3.67¡¾.76), patients (3.75¡¾.63) were above average. Satisfaction with IVIC300 of nurses differed significantly according to age, work unit, and clinical career. Satisfaction with IVIC300 of patients was not significantly different according to general characteristics.

Conclusion: These findings indicate that it is proven the accuracy of IVIC300, and is considered individual
characteristics in use IVIC300.