The Studies for the Malate Tissue Biosensor Using Malate Dehydrogenase(Decarboxylatin) in the Bundle Sheath Cell of the Corn Leaf




Hong Young-Seuk
Abstract
A biosensor for the measurement of malate has been constructed by the sodium-alginate immobilized bundle sheath cell tissue of corn leaf containing malate dehydrogenase(decarboxylating)(EC 1. 1. 1. 40) on the C02 gas-sensing electrode. The proposed tissue sensor had the linear in the range of malate concentration 5.6 x 10-,M-v2.5 x 10-¢¥M with a slope of 53.5 mV/decade in 0.02M Tris-HCI buffer solu¡©tion at optimum pH 8.0, and 25¡ÆC. A response time was 16. 18min. The present L-malate sensing tis¡©sue sensor is stable for more than one week. At pH 7.4, Km value was 0.6 x 10-¢¥M. The various kinds of salt did not effect the signal of malate tissue biosensor as the inhibitor. We can measure the malate by the C02 electrode at the pH=8.0. Thus, the proposed tissue sensor will be useful for the measure¡©ment of malate.