On-Line Measurement of Biochemical Oxygen Demand of Livestock Wastewater by Multi-Biosensor System
|
|
Kim Jin-Kyeung
Kim Tai-Jin Yang Chang-Bum
|
|
Abstract
|
|
|
The present study was intended to examine a basic scheme to determine the biochemical oxygen demand (BOD) of livestock wastewater by means of six individual dissolved oxygen (DO) sensors and its multi-measurable meter. Maximal point of the first order time derivative of the DO difference between DO distribution of sterilized livestock wastewater and that of non-sterilized livestock wastewater, was considered as the oxygen uptake rate (OUR) of microorganisms in livestock wastewater, as determined to be 0.00074 mg O2/§¤¡¤sec. The present study showed that there was a fair linear relationship (97.72%) between maximal OUR and BOD values of livestock wastewater, the latter being determined by classical Winkler azide method. It was thus concluded that the present multi-biosensor system might be applicable to an on-line system for measurement of BOD of livestock wastewater.
|
|
KEYWORD
|
|
Livestock, wastewater, BOD, OUR, monitoring
|
|