Korean Journal of Biotechnology and Bioengineering 2009 Volume.24 No. 3 p.253 ~ p.258
|
|
Optimization of Fabrication Conditions for Immunosensor Strip to Detect Escherichia coli O157 : H7
|
|
Park So-Jung
Kim Young-Kee
|
|
Abstract
|
|
|
In this study, the optimization of fabrication conditions was accomplished to make immuno-strip biosensor by the combination of enzyme linked immunosorbent assay (ELISA) and immuno-chromatographic strip techniques for the detection of Escherichia coli O157 : H7. Optimal fabrication conditions of capture antibody concentration, detection antibody concentration, and additive composition of running buffer solution were determined. Optimal concentration was determined as 1.0 mg/mL for both of capture antibody and detection antibody. A composition of 0.5% Tween20 and 3% BSA were selected as optimal additive for buffer solution to prevent non-specific binding.
|
|
KEYWORD
|
|
enzyme-linked immunosorbent assay (ELISA), Escherichia coli O157:H7, immunosensor, optimization
|
|