Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0380020150300040141
Korean Journal of Biotechnology and Bioengineering
2015 Volume.30 No. 4 p.141 ~ p.147
Development of a Validated Determination of Methylsulfonylmethane in Dietary Supplement by Gas Chromatography
Park Sang-Wook

Lee Won-Jae
Abstract
The convenient determination of methylsulfonylmethane (MSM) for a commercially available dietary supplement was developed using gas chromatography (GC)-flame ionization detector (FID). Chromatography was performed on a capillary column (0.32 mm I.D ¡¿ 30 m, 0.25 ¥ìm) coated with dimethylpolysiloxane using diethylene glycol methyl ether as an internal standard. The performance characteristics of GC were evaluated in terms of selectivity, linearity, precision, accuracy, recovery, limit of detection (LOD) and limit of quantification (LOQ). The calibration curve was highly linear (the coefficient of determination: 0.9979) within the concentration range of 10.0~800.0 ¥ìg/mL for MSM. The recoveries for three fortified concentrations were 96.7~97.1%, 96.6~97.3% and 96.8~97.2%, respectively. The LOD and LOQ of the method were 0.29 ¥ìg/mL and 0.97 ¥ìg/mL, respectively. All obtained results were acceptable according to the guidelines of the Association of Official Analytical Chemists for dietary supplements. Thus, the validated analytical method using the GC-FID system is suitable for the determination of MSM in dietary supplement formulations for quality control.
KEYWORD
Dietary supplement, Methylsulfonylmethane, Gas chromatography
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI)