Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0381919950250020080
Korean Journal of Microscopy
1995 Volume.25 No. 2 p.80 ~ p.87
Prerequisites on the Pre-installation and Installation of Analytical Electron Microscope
Kim Dae-Joong

Abstract
An analytical electron microscope system has been widely used in biology, medicine, veterinary medicine, agriculture, and materials, etc. nowadays in Korea Market since mid of 1980¡¯¡¯s. How to install and to choose the equipments? The answers are which prerequisites are needed to us. The purpose is going to introduce the prerequisites of the pre-installation and installation of Philips analytical electron microscope(CM 12/STEM and SEM 515, Philips, The Netherlands) in the National Institute of Safety Research, Seoul and to discuss the check-subjects. The check-subjects in the pre-installation and installation are more than 24. The influence of magnetic fields, mechanical vibrations, earth is crucial factor for decision of installation site. The areas of our electron microscope center are $105.6m^2$ and have the Automatic Image Analyzer System(IBAS, Kontron Co., Germany) connecting to the SEM mode. Water temperature was controlled by the NESLAB recirculatory chillers(NESLAB Co., U.S.A.).
KEYWORD
Pre-installation, Installation, Analytical Electron Microscope
FullTexts / Linksout information
 
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI)