KMID : 0381920150450020095
|
|
Korean Journal of Microscopy 2015 Volume.45 No. 2 p.95 ~ p.100
|
|
Temperature Calibration of a Specimen-heating Holder for Transmission Electron Microscopy
|
|
Kim Tae-Hoon
Bae Jee-Hwan Lee Jae-Wook Keesam Shin Lee Joon-Hwan Kim Mi-Yang Yang Cheol-Woong
|
|
Abstract
|
|
|
The in-situ heating transmission electron microscopy experiment allows us to observe the time- and temperature-dependent dynamic processes in nanoscale materials by examining the same specimen. The temperature, which is a major experimental parameter, must be measured accurately during in-situ heating experiments. Therefore, calibrating the thermocouple readout of the heating holder prior to the experiment is essential. The calibration can be performed using reference materials whose phase-transformation (melting, oxidation, reduction, etc.) temperatures are well-established. In this study, the calibration experiment was performed with four reference materials, i.e., pure Sn, Al-95 wt%Zn eutectic alloy, NiO/carbon nanotube composite, and pure Al, and the calibration curve and formula were obtained. The thermocouple readout of the holder used in this study provided a reliable temperature value with a relative error of <4%.
|
|
KEYWORD
|
|
Melting point, Dynamic observation, Annealing, Thermocouple, In-situtransmission electron microscopy
|
|
FullTexts / Linksout information
|
|
|
|
Listed journal information
|
|
|
|