Scatter Dose in soft tissue using the partial attenuation filter for 6 MY X-ray of linear accelerator


Abstract
Measured and calculated the TMR and SMR factors from percent depth dose underpartial attenuators which cover the whole part of the radiation beam with variousfilter thickness from 0 to 50 mm. This study was performed for x-ray beams generated with a 6 MV linear acceleratorat source to surface distance of 100cm in a water phantom for Lipowitz metal. TMR(0,d,t) was derived from non-linear polynomial regression with field sizedifferencies and a given filter thickness. In this experiments, the TMR(0,10,50) of 50mm of filter thickness was showed13.6 % higher than that of open field and SMR(5,10,50) was 38.5% smaller than thatof open field in same depth.
KEYWORD
FullTexts / Linksout information