KMID : 0384620060170040224
|
|
Korean Journal of Medical Physics 2006 Volume.17 No. 4 p.224 ~ p.231
|
|
Calculation of Energy Spectra for 6 MeV Electron Beam of LINAC Using MCNPX
|
|
Lee Jeong-Ok
Jeong Dong-Hyuk
|
|
Abstract
|
|
|
The electron energy spectra for 6 MeV electron beam were calculated using a MCNPX code. The head of the linear accelerator (ML6M; Mitsubishi, Japan) was modelled for this study. The energy spectrum of the initial electron beam was assumed to be Gaussian and the mean energy was determined by evaluating the measured and calculated values of and dose profiles in air. The energy distributions for electrons and photons at the interested points in the head of the linear accelerator were calculated by appling the Initial beam parameters. The effect of contaminant photons on depth dose curves were estimated by the photon energy spectra at the end of the applicator.
|
|
KEYWORD
|
|
MCNPX, Electron beam, Energy distribution
|
|
FullTexts / Linksout information
|
|
|
|
Listed journal information
|
|
|