Korean Journal of Medical Physics
2007 Volume.18 No. 3 p.139 ~ p.143
Dose Measurements using Phantoms for Tube Voltage, Tube Current, Slice Thickness in MDCT
Lee Chang-Re

Jeon Seong-Su
Nam So-Ra
Jo Hyu-Min
Jung Ji-Young
Kim Hee-Joung
Abstract
The purpose of this study was to measure and evaluate radiation dose for MDCT parameters. Patient dose for various combination of MDCT parameters were experimentally measured, using MDCT (GE light speed plus 4 slice, USA), model 2026C electrometer (RADICAL 2026C, USA), standard Polymethylmethacrylate (PMMA) head and body CT dosimetry phantoms. In clinical situations, for a typical abdominal scan performed with MDCT at 120 kVp, 180 mAs, 20 mm collimation, and a pitch of 0.75 CTDI_w,;CTDI_{vol} were measured as 20.2 mGy, 26.9 mGy, respectively. When scan length is assumed as 271.3 mm, DLP and measured effective dose of the abdominal would be calculated as 729.1;mGy{cdot}cm, 10.9 mSv, respectively.
KEYWORD
MDCT, Tube voltage, Tube current, Beam collimation, CTDI_w