Korean Journal of Medical Physics
2008 Volume.19 No. 1 p.56 ~ p.62
Study on the Dose Characteristics of the PTW-LinaCheck Dosimeter and Its Application to Daily Output Measurement
Jeong Dong-Hyeok

Lee Kang-Kyoo
Moon Un-Chul
Kim Hyeon-Jin
Kim Young-Seok
Moon Sung-Rock
Abstract
In this study, we have investigated the dose characteristics of PTW-LinaCheck designed to detect output of medical LINAC and discussed clinical use of the detector. The reproducibility, linearity, and dose rate dependency of the dosimeter were measured for photons of 6 and 15MV and the electrons of 4, 6, 9, 12, and 16MeV. To know the error ranges of the measured data in daily output measurement, the response variations due to geometrical setup errors were measured. As a result of measurement, the error range from the geometrical setup and the reproducibility was less than for given beam qualities in daily output measurement, where the errors from the linearity and the dose rate dependency were negligible. Finally, we concluded that the LinaCheck dosimeter has a good characteristics in terms of dose and setup convenience in daily output measurement. In addition we have shown an examples of clinical use of this dosimeter for measuring daily output more than 60 days.
KEYWORD
LinaCheck, Daily output, Setup error