Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0384620100210030261
Korean Journal of Medical Physics
2010 Volume.21 No. 3 p.261 ~ p.273
Signal and Noise Analysis of Indirect-Conversion Digital Radiography Detectors Using Linear-systems Transfer Theory
Yun Seung-Man

Lim Chang-Hwy
Han Jong-Chul
Joe Ok-La
Kim Jung-Min
Kim Ho-Kyung
Abstract
For the use of Indirect-conversion CMOS (complementary metal-oxide-semiconductor) detectors for digital x-ray radiography and their better designs, we have theoretically evaluated the spatial-frequency-dependent detective quantum efficiency (DQE) using the cascaded linear-systems transfer theory. In order to validate the developed model, the DQE was experimentally determined by the measured modulation-transfer function (MTF) and noise-power spectrum, and the estimated incident x-ray fluence under the mammography beam quality of W/Al. From the comparison between the theoretical and experimental DQEs, the overall tendencies were well agreed. Based on the developed model, we have investigated the DQEs values with respect to various design parameters of the CMOS x-ray detector such as phosphor quantum efficiency, Swank noise, photodiode quantum efficiency and the MTF of various scintillator screens. This theoretical approach is very useful tool for the understanding of the developed imaging systems as well as helpful for the better design or optimization for new development.
KEYWORD
Digital radiography, X-ray detector, Linear-systems transfer theory, MTF, NPS, DQE
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI) KoreaMed ´ëÇÑÀÇÇÐȸ ȸ¿ø