KMID : 0384620110220030148
Korean Journal of Medical Physics
2011 Volume.22 No. 3 p.148 ~ p.154
Study on the Evaluation of TRS-398 Quality Factors with Central Electrode Corrections for Small Cylindrical Chambers
Kang Yeong-Rok

Lee Chang-Yeol
Kim Jin-Ho
Moon Young-Min
Kwak Dong-Won
Kang Sang-Koo
Kim Jeung-Kee
Yang Kwang-Mo
Jeong Dong-Hyeok
Abstract
The quality factors (kQ,Q?) were evaluated by appling the results recently studied for the effect of central electrode in TRS-398 protocol. The PTW-31010 and IBA-CC13 chambers were used in this study. The quality factors were calculated as a function of beam quality for high energy electron and photon beams and compared with data currently used in TRS-398 protocol. In the PTW-31010 chamber using aluminium electrode, appling the new central electrode collections, the quality factors were 0.4% and 0.9% higher than current TRS-398 data for high energy photon and electron beams respectively. In the IBA-CC13 chamber using C-552 electrode, there are no variations in quality factors compared to TRS-398 data currently used
KEYWORD
TRS-398 protocol, Quality factor, Central electrode correction