KMID : 0384620120230020123
|
|
Korean Journal of Medical Physics 2012 Volume.23 No. 2 p.123 ~ p.126
|
|
Development of Prototype Quality Assurance Procedure for Blood Irradiator Using Glass Dosimeter Jig
|
|
Jeong Dong-Hyeok
Kang Yeong-Rok Moon Young-Min Kim Hyo-Jin Kwak Dong-Won Kim Jeung-Kee Lee Man-Woo
|
|
Abstract
|
|
|
For the purpose of quality assurance (QA) of the blood irradiator, QA programs for daily, monthly, and yearly were developed. For daily tests, simple items for basically operating the machine are recommended. For monthly and yearly tests, the measurement of dose to assure the dose delivery system are performed by a dosimetry devices (Glass dosimeter jig) developed in this study. The QA program is practical for clinical environment.
|
|
KEYWORD
|
|
Blood irradiator, Quality assurance, Glass dosimeter jig
|
|
FullTexts / Linksout information
|
|
|
|
Listed journal information
|
|
|