Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0384620120230040285
Korean Journal of Medical Physics
2012 Volume.23 No. 4 p.285 ~ p.291
The Measurement of National Standard -Rays Energy Spectrum
Kim Chul-Hang

Yi Chul-Young
Kim Hyun-Moon
Jeon Gook-Jin
Abstract
In the present study, we measured the pure beta particle energy spectra of 147Pm, 85Kr, 90Sr£«90Y radionuclide sources. We confirmed the residual maximum energies of KRISS sources meet the requirement of ISO 6980 and calculated mass collision stopping power ratio, which is essential for absolute measurement of absorbed dose from the reference ¥â-rays. The residual maximum energies of KRISS 147Pm, 85Kr, 90Sr£«90Y sources are 0.14, 0.57 and 0.93 MeV, respectively and the mass collision stopping power ratios are 1.123, 1.120 and 1.109, respectively.
KEYWORD
Beta-rays, Energy spectrum, The residual maximum energy, Mass collision stopping power ratio
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI) KoreaMed ´ëÇÑÀÇÇÐȸ ȸ¿ø