KMID : 0385520040170010082
|
|
Analytical Science & Technology 2004 Volume.17 No. 1 p.82 ~ p.84
|
|
The Adsorption Effect of Filter Paper on the Results of Trace Analysis
|
|
ÀÓÇ强/Lim HS
À̼®±Ù/Lee SG
|
|
Abstract
|
|
|
The quantitative trace analysis of ions could be deteriorated by filter papers because of the effect of adsorption. Generally the adsorption of anions on filter paper did not occurred. Instead, Cl^- and NO©ý^- ions were extracted from the filter papers. However, most metal ions were adsorbed on the filter papers by the formation of hydroxide in neutral solution. The adsorption of metal ions except Ag^+ion could be avoided depend on the acid concentrations.
|
|
KEYWORD
|
|
|
|
FullTexts / Linksout information
|
|
|
|