Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0385520050180060491
Analytical Science & Technology
2005 Volume.18 No. 6 p.491 ~ p.494
Potentiometric performances of polymer membrane electrode based on cyclosporin
Rhee IS
Abstract
The main component governing selectivity in ion-selective electrodes and optodes is the ionophore. For this reason, a member of natural products that possess selective ion-binding properties have long been sought after. By applying this principle, the performance of cyclosporin used as neutral carriers for calcium selective polymeric membrane electrode was investigated. The calcium ion-selective electrode based on cyclosporin gave a good Nernstian response of 26.6 §Æ per decade for calcium ion in the activity range 1¡¿10^(-6) M to 1¡¿10-©÷ M. The optimized calcium ion-selective electrode displayed very comparable selectivity for Ca©÷+ ion against alkali and alkaline earth metal ions, Na+, and Mg©÷+ in particular.
KEYWORD
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI)