KMID : 0385520080210060535
|
|
Analytical Science & Technology 2008 Volume.21 No. 6 p.535 ~ p.542
|
|
Analysis of clenbuterol in bovine muscle and milk by LC-ESI/MS/MS
|
|
Hong Se-Lyung
Jeong Ji-Yoon Park Hye-Jin Lee Soon-Ho Lee Jong-Ok
|
|
Abstract
|
|
|
A liquid chromatography-electrospray ionization tandem mass spectrometry (LC-ESI/MS/MS) method was developed for the determination and confirmation of clenbuterol in bovine muscle and milk. Clenbuterol and clenbuterol-D9 using as an internal standard in samples were extracted with ethyl acetate after hydrolysis and evaporated to dryness. The extracts were dissolved in 20% methanol and cleaned using HLB solid-phase extraction cartridge. The analytes were detected by LC-ESI/MS/MS on a C©û? column. Mass spectral acquisition was done in selected reaction monitoring (SRM) in positive ion mode to provide a high degree of sensitivity. Using MS/MS with SRM mode, the transitions (precursor to product) monitored were m/z 277¡æ203 for clenbuterol, and m/z 286¡æ204 for internal standard. The limits of quantitation (LOQ) and mean recoveries of clenbuterol in bovine muscle were 0.2 ¥ìg/kg and 84.3~91.1%, respectively. The LOQ and mean recoveries in milk were 0.05 §¶/§¸ and 87.7~98.3%, respectively.
|
|
KEYWORD
|
|
clenbuterol, LC-ESI/MS/MS, bovine muscle, milk
|
|
FullTexts / Linksout information
|
|
|
|
Listed journal information
|
|
|