KMID : 0385520090220020141
|
|
Analytical Science & Technology 2009 Volume.22 No. 2 p.141 ~ p.147
|
|
Investigation of the optimum condition for the quantitative analysis of Cu sample by Laser induced breakdown spectroscopy
|
|
Kim Seung-Hyun
Shin Hee-Sung Ju June-Sik Kim Ho-Dong
|
|
Abstract
|
|
|
A laser induced breakdown spectroscopy (LIBS) measurement was carried out to derive an optimized measurement condition with a high reproducibility and to grow a plasma sphere to 20 §® high under a 600 mtorr vacuum in order to improve an accuracy of measurement. The measurement of the plasma was taken at a 6.0 §® distance, in the direction of a plasma sphere, from a sample. This location belongs to the outer sphere region in the plasma. The calibration curve of ¡®Ni¡¯ and ¡®Cu¡¯ was acquired by the signal intensity ratio and the atomic ratio for the samples, and linear regression of ¡°Cu¡± was R©÷=0.9886, and the linear regression of ¡°Ni¡± was R©÷=0.9988. The accuracy of LIBS was improved pre-existence as the measurement error of ¡°Ni¡± was 0.78%.
|
|
KEYWORD
|
|
LIBS(laser induced breakdown spectroscopy), Quantitative analysis, Atomic spectroscopy, Analytical instruments
|
|
FullTexts / Linksout information
|
|
|
|
Listed journal information
|
|
|