Analytical Science & Technology 2011 Volume.24 No. 5 p.352 ~ p.359
|
|
Solvent-free determination of BTEX in water using repetitive membrane extraction followed by GC-MS
|
|
Kim He-Kap
Kim Se-young Lee Soo-Hyung
|
|
Abstract
|
|
|
An analytical method for solvent-free determination of benzene, toluene, ethylbenzene, and xylenes (BTEX) in water using repetitive membrane extractions coupled to cryofocusing and GC-MS was derived. BTEX compounds that permeated through a nonporous silicone membrane from the aqueous phase and evaporated into the acceptor phase were purged into a cryofocusing trap (-100 ¡É) with helium gas. The BTEX compounds, thus enriched in the trap, were thermally desorbed into a capillary column GC and detected using an MS. The flow rate of the donor phase (30 mL water) was set at 10 mL/min, and membrane extractions, accomplished by returning the water drained from the extraction module to the sample container, were repeated three times at 20 ¡¾ 2 ¡É. Although recoveries (%) were variable, from the highest for benzene (approximately 80%) to the lowest for ethylbenzene and xylenes (3.5-10%), the method showed satisfactory precision (RSD 2.2-10%) with good-linearity calibration curves (r©÷ 0.9976-0.9997 in 1-100 §¶/L range) for all of the compounds. The method detection limits (MDLs) ranged from 0.16 to 1.8 §¶/L. The results showed the method¡¯s advantages such as short analysis time and overall simplicity without solvent compared to the conventional techniques.
|
|
KEYWORD
|
|
BTEX, cryofocusing, GC-MS, repetitive membrane extractions, nonporous silicone membrane
|
|