Abstract
Optical purities of 10 commercialized naproxens prepared from eight Korean drug companies were examined by an optimized chiral HPLC condition. The Chiralcel OD-H column and ChiralHyun-LE(S)-1 column were used as chiral stationary phases and the mixed eluent of hexane/isopropanol/acetic acid as 100:2.85:0.1 was used as a mobile phase for effective enantioseparation. Optical purity values of most samples were higher than 97 percents, only one of them was about 95 percents. The average relative standard deviation of them appeared very small (0.034%).
KEYWORD
chiral chromatography, optical purity, commerical naproxen