KMID : 0385520120250010060
|
|
Analytical Science & Technology 2012 Volume.25 No. 1 p.60 ~ p.68
|
|
Trace level analysis of 25 semi-volatile organic compounds in surface water by gas chromatography-mass spectrometry
|
|
Kim Tae-Seung
Hong Suk-Young Kim Jong-Eun Oh Jin-A Shin Ho-Sang
|
|
Abstract
|
|
|
A gas chromatography-mass spectrometric (GC-MS) method was developed for determining 25 semivolatile organic compounds in water. A 1.0 L water sample was placed in a separatory funnel and saturated with NaCl, and the solution was extracted two times with 40 mL of methylene chloride. Under the established condition, the linear quantification range was 0.02-800 ng/L and the relative standard deviation was less than 15%. The method was used to analyze 16 surface water samples collected from various regions in Gum-River. The samples revealed SVOC concentrations in the range of 0.02-96.8 ng/L. Maximum concentrations of VOCs detected were not exceeded the EPA or Germany guidelines in any of the samples. The developed method may be valuable for monitoring SVOCs in water.
|
|
KEYWORD
|
|
semi-volatile organic compounds, gas chromatography-mass spectrometry, surface water,
|
|
FullTexts / Linksout information
|
|
|
|
Listed journal information
|
|
|