KMID : 0385520120250030190
Analytical Science & Technology
2012 Volume.25 No. 3 p.190 ~ p.196
Trace level analysis of Pb in plasma by inductively coupled plasma/mass spectrometry
Lee Sung-Bae

Yang Jeong-Sun
Choi Sung-Bong
Shin Ho-Sang
Abstract
The human exposure of lead has usually detected the amount of lead in the whole blood, however, this method has a shortcoming to give the information on the short-term exposure to lead. In that sense, it is desirable to estimates the level of lead in plasma to draw the chronic bio-marker of lead exposure even though it is difficult to measure lead of several ng/L. An inductively coupled plasma-mass spectrometry (ICPMS) method was developed for determining lead in plasma as the chronic bio-marker of lead of workers. To minimize the contamination of lead from the environment, we constructed class 1,000 clean room and compared the amount of floating dust before and after the operation of the clean room. The limit of detection (LOD) and the limit of quantification (LOQ) of lead in fetal bovine serum were 4.3 ng/L and 12.2 ng/L by NIOSH method (statistical calculation method) and 7.0 ng/L and 22.1 ng/L by signal/noise ratio, respectively. The
accuracy was in a range of 92.3-101.3%, and the precision of the assay was less than 4% in the samples spiked in the concentration of 20 ng/L and 2,000 ng/L. The method was simple, reproducible and sensitive enough to permit reliable analysis of lead to the ng/L level in plasma and/or serum. The method was also useful for the biological monitoring of chronic exposure to lead.
KEYWORD
lead, plasma, inductively coupled plasma-mass spectrometry, chronic bio-marker
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI)