KMID : 0385520130260040221
|
|
Analytical Science & Technology 2013 Volume.26 No. 4 p.221 ~ p.227
|
|
Validation of LC-MS/MS method for determination of ginsenoside Rg1 in human plasma
|
|
Kim Yun-Jeong
Han Song-Hee Jeon Ji-Young Hwang Min-ho Im Yong-Jin Lee Sun-Young Chae Soo-Wan Kim Min-Gul
|
|
Abstract
|
|
|
A sensitive and selective liquid chromatography-tandem mass spectrometry (LC-MS/MS) was developed for the investigation of the ginsenoside Rg1 in human plasma. After addition of internal standard (digoxin), plasma was diluted with acetone and methanol (80:20), the supernatant was concentrated and analyzed by LC-MS/MS. The optimal chromatographic separation was achieved on an Agilent Eclipse XDB-C18 column (4.6 ¡¿ 150 mm, 5 ¥ìm) with a mobile phase of 0.1% formic acid in water and 0.1% formic acid in methanol at a flow rate of 0.9 mL/min gradient mode. The standard calibration curve for ginsenoside Rg1 was linear (r 2 =0.9995) over the concentration range 1~500 ng/mL in human plasma. The intra- and inter-day precision over the concentration range of ginsenoside Rg1 was lower than 7.53% (correlation of variance, CV), and accuracy exceeded 98.28%. This LC-MS/MS assay of ginsenoside Rg1 in human plasma is applicable for quantifying in the pharmacokinetic study.
|
|
KEYWORD
|
|
ginsenoside Rg1, LC-MS/MS, validation, human plasma,
|
|
FullTexts / Linksout information
|
|
|
|