KMID : 0385520140270030161
|
|
Analytical Science & Technology 2014 Volume.27 No. 3 p.161 ~ p.166
|
|
Development of alternative assay method without hazardous reagent, carbon tetrachloride, for cefroxadine capsules in Korean Pharmaceutical Codex
|
|
Seo Na-Rae
Kang Yeo-Ul Park Yeong-Shin Lee Jong-Min Jeong Myung-Jin Jeon Sang-Seol An Jae-Hyung Kim Kyeong-Ho
|
|
Abstract
|
|
|
Currently oxidation-reduction titration method for the assay method using the hazardous reagent, carbon tetrachloride, for cefroxadine capsules has been used in Korean Pharmaceutical Codex. We developed and validated an alternative HPLC assay method by substituting the use of the hazardous reagent, carbon tetrachloride, to the use of less toxic ones like acetonitrile. The linearity of the calibration curves in the desired concentration range was good (r2 > 0.999). Relative standard deviations of intra-day precision were ranged between 0.30% and 0.69% and inter-day precision were ranged between 0.47% and 0.82%. Accuracy was obtained with recoveries in range of 100.20% and 100.56%. Developed method was applied to the determination of cefroxadine in commercial cefroxadine capsule and could be expected to become valuable tools for revising the Korean Pharmaceutical Codex.
|
|
KEYWORD
|
|
Korean Pharmaceutical Codex, hazardous reagents, alternative testing methods, validation, high performance liquid chromatography,
|
|