Validation of analysis of urinary fluoride by ion selective electrode method
Lee Mi-Young

Yoo Kye-Mook
Abstract
A simple and sensitive analytical method for fluoride in urine by ion selective electrode (ISE) method was presented. Traditional buffer for fluoride determination using ISE is acetate-based one. Researchers have pointed out some drawbacks of the buffer for fluoride ISE analysis, and some other buffers including citrateammonium buffer and MES buffer have been studied for accurate determination of fluoride in urine here. These buffers provided promising results in environmental field, and this author focused on overcoming the interference of co-existing aluminium. The results show that MES-CyDTA buffer gave the best recovery with accuracy of 95-97.5% and precision of 1.9-7.9% for reference sample of 1.8-7.8 mg/L fluoride in urine, with smaller amount of samples and shorter analysis time compared with the traditional method which used acetate buffer. The method was applied to field samples, and which showed urinary of 0.98¡¾0.38 mg/g creatinine for workers in electric cable manufacturing factory (n=15) and 0.59¡¾0.30 mg/g creatinine for non-exposed workers (n=12).
KEYWORD
urinary fluoride, biological monitoring, ion selective electrode, MES-CyDTA buffer,