Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0385520180310010023
Analytical Science & Technology
2018 Volume.31 No. 1 p.23 ~ p.30
Simultaneous determination of aromatic material causing allergic in children¡¯s products by Gas Chromatography
Ko Kyeong-Mok

Rhu Chan-Joo
Kim Jong-Won
Lee Seok-Ki
Abstract
Twenty-two allergy-induced aromatics in children were analyzed using a gas chromatography flame ionization detector (GC-FID) and gas chromatography mass spectrometer (GC-MSD). Analytes were extracted using an automatic Soxhlet extractor and centrifuged for 10 min in a fast freezing centrifuge, and the supernatant was transferred into a 2 mL vial and injected in split mode. Under the established conditions, the calibration curve showed linearity with a correlation coefficient of 0.996 or more. A wide range of sensitivity of 6.7 to 1,859,839 depending on the device characteristics and detector used was shown. The detection limit of the device was 0.0032 to 0.0335 ¥ìg/mL, and the maximum detection limit was less than 0.1 ¥ìg/mL. The detection limit of the method ranged from 0.0033 to 0.1161 ¥ìg/mL. In addition, the limit of quantification ranged from 0.0100 to 0.5422 ¥ìg/mL, with a level of precision ranging from 0.21 % to 4.89 % and a degree of accuracy ranging from 89 % to 111 %. The analytical method developed in this study was applied to commercial products.
KEYWORD
allergy, aromatic material, children¡¯s products, GC-FID, GC-MSD
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI)