Analytical Science & Technology
2019 Volume.32 No. 1 p.24 ~ p.34
analysis of 16 sedative-hypnotics and sleep-inducers by ultra-high-performance liquid chromatography with UV detection (UPLC-UV) and quadruple Orbitrap mass spectrometry (Q-Orbitrap-MS)
Lee Ji-Hyun

Choi Ji-Yeon
Park Han-Na
Min Ah-Young
Kim Nam-Sook
Park Seong-Soo
Park Sung-Kwan
Kang Ho-Il
Abstract
The safety of food is occasionally questionable, as there have been some reports of products contaminated with illegal adulterants. In this study, the presence of 16 sedative-hypnotics and sleep inducers in dietary supplements was determined by ultra-high-performance liquid chromatography with UV detection (UPLC-UV) and quadruple Orbitrap mass spectrometry (Q-Orbitrap-MS). The UPLC method was validated, providing a linearity (R2) of more than 0.999, and LODs and LOQs that ranged from 0.2 to 0.5 and 0.6 to 1.5 ¥ìg mL?1, respectively. The repeatabilities were 0.2-8.4 % (intra-day) and 0.3-4.5 % (inter-day), and the accuracies were 89.0-117.0 % (intra-day) and 87.8-111.9 % (inter-day). The mean recoveries of the spiked samples ranged from 98.7 to 107.3 %. The relative standard deviation (%RSD) of the stability was less than 2.4 %. Using the developed method, one sedative-hypnotic compound, phenobarbital, was detected in one of the nineteen samples tested. In addition, the major characteristic fragment ions of each target compound were confirmed using Q-Orbitrap-MS for higher accuracy. Monitoring the presence of these 16 sedative-hypnotics and sleep inducers in dietary supplements should be pursued in the interest of human health, and the results of this study confirmed that the developed method has value for this application.
KEYWORD
Sedative-Hypnotics, Adulterants, UPLC-UV, Q-Orbitrap-MS, Method Validation
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI)