Analytical Science & Technology
2020 Volume.33 No. 2 p.68 ~ p.75
Development of nitrogen and oxygen certified reference materials in 10 ¥ìmol/mol for the purity evaluation
Ahn Byung-Soo

Moon Dong-Min
Lee Jin-Bok
Kim Jin-Seog
Lee Jin-Hong
Hong Ki-Ryong
Abstract
We have developed 10 ¥ìmol/mol nitrogen and oxygen certified reference materials (CRMs) in helium, as a SI-traceable gas standard for a quantifying of impurities in pure gases for the first time in Korea. The standard gas mixtures of nitrogen and oxygen were prepared in 5000 ¥ìmol/mol and sequentially were diluted to 250 ¥ìmol/mol and 10 ¥ìmol/mol according to the gravimetric preparation. In each dilution step, two cylinders of CRMs were prepared. The verification of internal consistency among the prepared gas mixtures was performed by using GC-TCD. The amount fractions and those expanded uncertainties (k = 2) of nitrogen and oxygen in the standard gas mixtures were (10.12 ¡¾ 0.08) ¥ìmol/mol and (10.18 ¡¾ 0.08) ¥ìmol/mol for nitrogen, and (9.88 ¡¾ 0.06) ¥ìmol/mol and (9.94 ¡¾ 0.06) ¥ìmol/mol for oxygen, respectively. We have conducted a purity assessment of two commercial helium gases using developed CRMs. As the results of the purity assessment, nitrogen and oxygen were detected by (1.66 ¡¾ 0.03) ¥ìmol/mol and (0.31 ¡¾ 0.02) ¥ìmol/mol, respectively, as the impurities in one of the pure helium.
KEYWORD
gas analysis, purity assessment, gas chromatography, thermal conductivity detector, certified reference material