Analytical Science & Technology 2022 Volume.35 No. 6 p.249 ~ p.255
|
|
Development of analysis method for high purity nitrogen using GC-FID/Methanizer
|
|
You Je-I
Lee Jin-Bok Kim Jin-Seog Kim Woon-Jung Hong Ki-Ryong
|
|
Abstract
|
|
|
In this study, a new method for the analysis of high-purity nitrogen was developed. A gas chromatography-flame ionization detector (GC-FID) was used for purity analysis. Certified reference materials (CRMs) at a level of 3 ¥ìmol/mol of carbon monoxide (CO), carbon dioxide (CO2), and methane (CH4), which may exist in high-purity nitrogen, were prepared using the gravimetric method, and these CRMs were used for purity analysis. In this new method, ultra-high-purity and high-purity nitrogen were used as carrier gases. The impurities in high-purity nitrogen were quantitatively analyzed by comparing the differences in the area values of the GC chromatograms of the prepared CRMs. We purchased liquid nitrogen and three bottles of nitrogen gas, which were produced by three different manufacturers, using high-purity nitrogen. Furthermore, to validate the developed purity analysis method, the fraction of impurities in high-purity nitrogen was compared with the results of the typical purity analysis method. The comparison results were consistent within the expanded uncertainties (k = 2).
|
|
KEYWORD
|
|
gas chromatography, certified reference material (CRM), gravimetric preparation method, purity analysis, high purity nitrogen
|
|