KMID : 0386420070200010070
|
|
Journal of the Korean Fracture Society 2007 Volume.20 No. 1 p.70 ~ p.75
|
|
Biomechanical Efficacy of Various Anterior Spinal Fixation in Treatment of Thoraco-Lumbar Spine Fracture
|
|
Park Ye-Soo
Kim Hyoung-Jin Choi Choong-Hyeok Park Won-Man Kim Yoon-Hyuk
|
|
Abstract
|
|
|
Purpose: To evaluate the biomechanical results according to various anterior spinal fixation methodology in the treatment of thoracolumbar spine fracture.
Materials and Methods: The comparative analysis of fixation method was evaluated by three dimensional finite element model using the 1 mm reconstruction image of CT. Authors evaluated the flexion, extension, lateral bending, torsional stresses with 12 fixation methods for the compression and burst fracture.
Results: In biomechanical analysis, stiffness of body-fixation device was more stable in two-rod system in compression fracture and was stable in one-rod, two-rod system in burst fracture, but two-rod system was showed over-increase of stiffness.
Conclusion: Authors recommend the usage of two-rod system in anterior fixation only and anterior one-rod system in anterior-posterior fixation.
|
|
KEYWORD
|
|
Thoracolumbar spine fracture, Anterior fixation, One-rod system
|
|
FullTexts / Linksout information
|
|
|
|
Listed journal information
|
|
|