KMID : 0386420120250020136
|
|
Journal of the Korean Fracture Society 2012 Volume.25 No. 2 p.136 ~ p.141
|
|
Repeated Metal Breakage in a Femoral Shaft Fracture with Lateral Bowing -A Case Report-
|
|
Kim Dong-Soo
Kim Yong-Min Choi Eui-Sung Shon Hyun-Chul Park Kyoung-Jin Cho Byung-Ki Park Ji-Kang Lee Hyun-Cheol Hong Kyung-Ho
|
|
Abstract
|
|
|
Fractures of the femoral shaft with marked bowing face some obstacles in fixation of the fracture such as difficulty in insertion of the intramedullary nail (IM nail) or exact contouring plate. Locking compression plates (LCP) are an option to manage this problem. However, we experienced consecutive breakage of LCP twice and IM nail once in an 80-year-old female. Finally, union of the fracture was achieved after fixation of the IM nail and additional plate together. Fractures of the femur shaft with marked bowing are thought to have different biomechanical properties; therefore, we present this case with a review of the literature.
|
|
KEYWORD
|
|
Femur shaft fracture, Lateral bowing, Metal breakage, Locking compression plate, Intramedullary nail
|
|
FullTexts / Linksout information
|
|
|
|
Listed journal information
|
|
|