Breakage of Reamer during Tibia Intramedullary Nailing
|
|
Kwak Ho-Yoon
Kim Chin-Soo Young Ki-Won Joh Joo-Won Hwang Sae-Min
|
|
Abstract
|
|
|
The reamer crack, followed by breakage at its distal part occurred during intramedullary nailing of tibial shaft fracture. The broken reamer was trapped in the intramedullary canal, making it very difficult to pull out. We successfully extracted the broken reamer by retrograde impaction through the fracture site and completed intramedullary nailing procedure. Thus, we present this case with a review of the literature.
|
|
KEYWORD
|
|
Tibia, Fracture, Intramedullary nailing
|
|