KMID : 0386720060180050507
|
|
Journal of the Korean Hip Society 2006 Volume.18 No. 5 p.507 ~ p.511
|
|
Disengagement of the Lag Screw from Side plate in a Subtrochanteric Fracture of the Femur treated with Compression Hip Screw - A Case Report -
|
|
Shon Hyun-Chul
Kim Yong-Min Kim Dong-Soo Choi Eui-Sung Park Kyoung-Jin Kim Seok-Won
|
|
Abstract
|
|
|
Subtrochanteric fractures of the femur can be treated by sliding hip screw devices. We encountered a subtrochanteric fracture in a 43 years old male. Open reduction and fixation with a sliding hip screw and plate were performed. One week later, the sliding screw disengaged from the side plate. In the operating room, the side plate was changed to the long barrel type and a compression screw was inserted. At two year follow-up, the fracture was united with good function.
|
|
KEYWORD
|
|
Subtrochanteric fracture of femur, Disengagement, sliding hip screw
|
|
FullTexts / Linksout information
|
|
|
|
Listed journal information
|
|
|