Abstract
|
|
|
Purpose:To evaluate the minimum 10-year follow up results of primary total hip arthroplasty (THA) performed using a sandwich-type (alumina-polyethylene-titanium) ceramic bearing.
Materials and Methods : Thirty four patients (40 hips) who underwent a THA with sandwich typed liners from November 1998 to December 2000 were analyzed. Among the 34 patients, 25 were men and nine were women. Mean follow-up was 134 months (range, 120~145 months) and mean patient age at the time of THA was 47.1 years (range, 24~65 years). The clinical results were evaluated using the Harris hip score and level of thigh pain with a limping gait. The radiographic evaluation was done in terms of the endosteal new bone formation, radiolucent line, subsidence of the stem, migration of the acetabular cup, and proof of loosening.
Results:The mean preoperative Harris hip score of 50.2 points (range, 31~87 points) was improved to 90.9 points (range, 75~99 points) at the final follow-up and thigh pain with limping gait in one case. All cases had fixation by bony ingrowth. No radiographically detectable loosening was observed in any hip. Fracture of ceramic liner in one case required change of the polyethylene liner.
Conclusion:At the minimum 10-year follow up, survival rate as the end-point was favorable. However, fracture of the ceramic is still a major problem. Satisfactory results can be obtained by more precise surgical technique for acerabular inclincation and anteversion, and by improving the quality of the ceramic.
|
|