Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0387719960070010049
Korean Journal of Blood Transfusion
1996 Volume.7 No. 1 p.49 ~ p.55
CS-3000 Plus Cell Separator : 4-35 collection Chamber vs PLT-30 Collection Chamber
À̽¹«
½ÅÈñºÀ/³²Á¤Çö/±èÇö¿Á/±Ç¿ÀÇö
Abstract
Background:
recently, it has been established that plateletpheresis needs more efficiency and shorter ocessing time. Fenwall laboratories developed a new collection chamber for CS-3000 Plus, PLT-30TM ollection chamber, which can reduce the processing time
with
efficient collection, We evaluated the LT-30TM collection chamber by comparing it with A-35 collection chamber that has been used as a andard collection chamber of CS-3000 Plus. Us
methods:
Thirty platelet collection procedures were performed using the CS 3000 Plus with A-35 ollection chamber and PLT-30TM collection chamber. The changes of the hematologic parameters tween pre-and post-donation in donors and the total platelets
yields
and the contaminated WBCs in the ateletpheresis products were evaluated. In processing, the yield predictor calibration was adjusted to 90 and 1.13 in A-35 and PLT-30TM respectively. Yield predictors of pheresis were the same as 3.5¡¿10E11 both
and
end
point volumes were calculated from the CS-3000 Plus. Processing volume and ocessing times were compared between A-35 and PLT-30TM groups.
Results:
With PLT-30TM collection chamber 3.38¡¾0.72¡¿10E11/L platelets were harvested, whereas 3.20¡¾73¡¿10E11/L were collected with A-35 collection chamber. Which was not significantly different. But ocessing time with the PLT-30TM collection chamber
was
more reduced than that with the A-35 collection amber by about 20 minutes (PLT-30TM: 88.6¡¾8.4 min, A-35: 106.7¡¾11.7min). Collection efficiency of T-30TM chamber was 50.7¡¾12.5% and that of A-35 chamber was 44.4¡¾8.8%. The leukocyte
contamination
the
platelet concentrates were not statistically different (PLT-30TM: 0.0-3.6¡¿10E6, A-35: 0.1-4.1¡¿10E6).
Conclusions:
PLT-30TM collection chamber has the advantages of shortening the donation time and decreasing the processing volume with better collection efficiency and flexibility of Platelet concentrate volume. (Korean J Blood Transfus 7(1): 49~55, 1996)
KEYWORD
FullTexts / Linksout information
 
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI) KoreaMed ´ëÇÑÀÇÇÐȸ ȸ¿ø