KMID : 0387720060170010039
|
|
Korean Journal of Blood Transfusion 2006 Volume.17 No. 1 p.39 ~ p.47
|
|
Comparison of Amicus and CS-3000 Plus for the Collection of Peripheral Blood Stem Cells.
|
|
Lee Jin-Sol
Cho Duck Shin Myung-Geun Ryang Dong-Wook Lee Je-Jung Kim Hyeoung-Joon Chae Hong-Jae Shin Jong-Hee Suh Soon-Pal
|
|
Abstract
|
|
|
Background: The efficient collection of peripheral blood stem cells (PBSC) from donors who donate for allogeneic transplants as well as from patients undergoing autologous transplants is essential for a successful transplant. Recently, the Amicus cell separator and the associated MNC collection computer software program for PBSC collection were introduced in Korea.
Methods: Two apheresis machines (Amicus, Baxter Healthcare; and CS-3000 plus, Baxter Healthcare) were compared retrospectively. A total number of 144 procedures were performed on 14 donors and 28 patients. The pre- and post-apheresis complete blood cell (CBC) counts and the number of hematopoietic progenitor cells (HPC) were determined in the peripheral blood from the subjects. The CBC, HPC, CD34+ cell counts and the level of colony-forming unit-granulocyte-macrophages (CFU-GM) were measured in the PBSC product collected from both machines.
Results: Both machines collected a similar number of CD34+ cells from the donors and patients. On the other hand, the Amicus collected significantly more nucleated cells, MNCs, HPCs and CFU-GM in the patients with significantly less RBC contamination than those with CS-3000 plus. The decrease in the peripheral blood platelet counts in the donors and patients was more prominent after apheresis using the CS-3000 plus (117.00+/-42.75 x 10(3)/microliter, 61.22+/-43.62 x 10(3)/microliter) than Amicus (26.04+/-18.68 x 10(3)/microliter, 22.15+/-28.66 x 10(3)/microliter)(p<0.05).
Conclusion: PBSC collection can be performed successfully using CS-3000 plus and Amicus. Amicus is superior to CS-3000 plus in avoiding apheresis-induced thrombocytopenia, and is expected to prevent unnecessary platelet transfusion.
|
|
KEYWORD
|
|
Peripheral blood stem cell collection, Amicus, CS-3000 plus, Platelet contaminaion
|
|
FullTexts / Linksout information
|
|
|
|
Listed journal information
|
|
|
|