Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0390619990070020148
Journal of Cardiovascular Ultrasound
1999 Volume.7 No. 2 p.148 ~ p.155
Real Time Determination of Left Atrial Volume by Automatic Boundary Detection: Comparison with Two-dimensional Echocardiographic Measurement
Ha Jong-Won

Kang Seok-Min
Rim Se-Joong
Kim Myung-Gon
Goh Choong-Won
Kim Ji-Young
Oh Eun-Kyung
Hwang Eun-Kyung
Kim Hyun-Joo
Chung Nam-Sik
Abstract
Background£ºEchocardiographic automatic boundary detection (ABD) is a new on-line technique which automatically outlines the left atrial endocardial border and instantly calculates the left atrial area and volume from two dimensional echocardiographic images. To determine if left atrial volume can be derived using the ABD method, 70 patients with various etiology of heart disease (mean age 51) underwent complete echocardiographic examination with both the ABD method and conventional two-dimensional echocardiographic measurement.

Methods£ºEnd-systolic left atrial volume was obtained on-line from the apical four chamber view using ABD (ABD LAVOL). Left atrial length was also measured from an apical view. Left atrial volume was calculated using
ellipsoid model (2D LAVOL) and method of disc (MANUAL LAVOL) by two-dimensional echocardiography.

Results£ºThe end-systolic left atrial volume by ellipsoid model, method of disc and ABD were 87¡¾31ml, 57¡¾25 and 60¡¾28ml, respectively. Linear regression analysis revealed that the ABD LAVOL is closely correlated with 2D LAVOL and MANUAL LAVOL (ABD LAVOL vs 2D LAVOL, SEE£½22ml, r£½0.65, p<0.001£»ABD LAVOL vs MANUAL LAVOL, SEE£½6ml, r£½0.97, p<0.001).

Conclusion£ºABD echocar-diography is a new on-line technique which may be used to accurately calculate left atrial volume in patients with various heart diseases.
KEYWORD
Left atrial volume, Automatic boundary detection
FullTexts / Linksout information
 
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI) KoreaMed ´ëÇÑÀÇÇÐȸ ȸ¿ø