Àá½Ã¸¸ ±â´Ù·Á ÁÖ¼¼¿ä. ·ÎµùÁßÀÔ´Ï´Ù.
KMID : 0604519880140010001
Journal of the Society of Cosmetic Scientists of Korea
1988 Volume.14 No. 1 p.1 ~ p.15
DETERMINATION OF TRIETHANOLAMINE BY HIGH PERFORMANCE LIQUID CHROMATOGRAPHY WITH POST COLUMN REACTION
Kim Jin-Woo

Kim Seung-Jung
Lee Bo-Seob
Abstract
A new method for liquid chromatography with post column reaction is suggested for the separation and quantification of tertiary amines. A mixture of triethanolamine and N-ethyl diethanolamine was separated by a strong cation exchange column, followed by spectrophtometric detection of the blue colors generated from the reaction of each amine with the Folin-Ciocalteau reagent. The tertiary amines were properly separated when an eluent of pH 9.5 containing 0.5M sodium nitrate was used. Under this condition, calibration curve of triethanolamine in 2-10mg/100ml concentration range was attained. Good results were obtained when cream and shampoo preparations containing known amount of triethanolamine were analysed according to this method. In case the sample did not contain any other interfering reducing substances, the amine was quantitatively determined by the simple reaction of the samples with Folin-Ciocalteau reagent, and the subsequent spectrophotometric measurement.
KEYWORD
FullTexts / Linksout information
Listed journal information
ÇмúÁøÈïÀç´Ü(KCI)